Insert the electrode into the nib and fix the electrode to the nib.Connect the electrosurgical pencil to the electrosurgical unit's 3-pin connector, and then use it according to the electrosurgical unit's boot-up procedure.The electrosurgical pencil with electrodes is a device that uses high-frequency current to remove tissue and control bleeding.Depending on the surgical site, the surgeon should choose the electrode best suited to the procedure.
Scope of application: Operating room; surgical operator.
Product Specifications:
Pencil size:
| 149×12.24×12.69㎜
|
Cable length:
| 3m Silicone
|
Control method:
| Hand control
|
Button design:
| Round button
|
Pencil material:
| Polysulfone
|
Connector type:
| 4mm banana plugs (3pcs) Three-plug foolproof design
|
Technical symbols:Yellow is the cutting function; Blue is the coagulation function.
Cleaning, disinfection and preventive inspection:
- Cleaning: Wipe the components with a sponge or cloth soaked in 0.1% Teepol cleaning solution, then rinse with clean water and wipe dry.
- Disinfection: The components must be cleaned according to the cleaning procedures described above. Then wipe with a sponge or cloth soaked in 75% industrial denatured alcohol solution.
- Sterilization: Before carrying out any sterilization process, you must remove any carbonization from the electrode and replace the electrode within the electrosurgical pencil socket.
- Steam high temperature and high pressure: Set temperature to 132°C for 20 minutes.